Call: 734.761.1270E-mail: info@enmet.com
All Categories > Sensor/Transmitters, Controllers, Gas Samplers, Single-Point Monitors, Specialty Monitors > IR-6000 Infrared Hydrocarbon Sensor/Transmitter > IR-6000 Infrared Hydrocarbon Sensor/Transmitter > IR-6000 Infrared Hydrocarbon Sensor/Transmitter  
Request Information

IR-6000

larger image

IR-6000 Infrared Hydrocarbon Sensor/Transmitter



The IR-6000 hydrocarbon detector is a single source dual wavelength instrument. The sensing and reference elements are self-compensating for optical integrity and other signal inhibitors. The 4-20 mA analog output can be connected to one of the several MX-Series Controllers available from ENMET. These controllers furnish 24 Vdc power for sensor/transmitters and provide a digital display of gas concentration, audio/visual alarm and relays.

Click here for other Sensor/Transmitters Available.

Click here for Controllers Available.

Click here for Gas Samplers Available.




Features · Applications · Specifications

Features
  • Infrared sensing technology
  • Low maintenance
  • Immune to poisoning
  • Designed for harsh environments
  • Explosionproof
  • Fast response time
  • Self-compensating optical bench
  • No moving parts
  • Heated optical chamber
  • Low power consumption
  • Operates in constant hydrocarbon background
  • Operates in anaerobic atmospheres
  • 4 to 20 mA output
  • 0 to 100% LEL detection range
  • Digital display option available
  • Stainless steel construction



Applications
The IR-6000 hydrocarbon detector is designed to be used in the same applications where catalytic bead type combustible gas sensors have been applied. A partial list of example applications would include:
  • Refineries
  • Gas turbines
  • Drilling and production platform
  • Chemical plants
  • Fuel loading facilities
  • Compressor stations
  • Oil well logging
  • Wastewater treatment facilities
  • LNG/LPG processing and storage facilities
  • Transportation facilities



Specifications

Operating Voltage

24 Vdc (18-32 Vdc)

Current Draw

Average: 210 mA
Peak: 400 mA

Power Consumption

Max. 5 W

Analog Output

4-20 mA

Range

0-100% LEL

Standard Calibration

Methane, Propane and Butane

Other Available Calibrations

Hexane, Isobutane, Ethane, Octane, Ethanol, Ethylene, Benzene, Methanol, Ethylene oxide, Ethyl benzene, 1-Butene, Toluene, Cyclohexanone, Propylene, Xylene, Cyclohexanol, n-Butane, Heptane, Dimethylpropane. The IR-6000 can be calibrated for most hydrocarbons, provided a calibration gas is available.

Accuracy

± 3% 0 to 50% LEL
± 5% 51 to 100% LEL

Response Time

T50 < 5 seconds
T90 < 10 seconds

Temperature Range

-40 to +158 °F (-40 to +70 °C)

Humidity Range

0-99% (non-condensing)

Housing Construction

316 Stainless Steel.

Approval

CSA

Classification

Class 1, Div. 1, Groups B, C & D

Ingress Protection

IP 54

Mounting

3/4" NPT

Size

6.5 x 2.6 in (165 x 66 mm)

Weight

2 lbs (0.9 kg)


    Download PDF Download PDF  
    Email Email This Page  
    Save To Favorites Save To Favorites
  

 DOWNLOADS 

 width= IR-6000 Manual
(PDF, 114KB)

 width= IR-6000 Literature
(PDF, 66KB)


Request Information
Site created by ThomasNet Web Solutions and powered by Navigator Platform
ENMET Corporation | Gas Detection for Health, Safety and Process Monitoring