Call: 734.761.1270E-mail: info@enmet.com
All Categories > Personal, Single Gas Detectors > Glutaraldemeter®3 > P/N item-1114  
Request Information

Glutaraldemeter

larger image

P/N item-1114, Glutaraldemeter®3



The user friendly Glutaraldemeter® 3 from PPM Technology Ltd. gives a rapid indication of whether the glutaraldehyde level is above or below the Maximum Exposure Level.
  • Uses the well proven Lion electrochemical fuel cell sensor
  • Has improved selectivity
  • Stores time to peak readings
  • Holds peak reading
  • Stores last ten values
The GLUTARALDEMETER® 3 incorporates the latest technology in its design. Although it uses the same fuel cell and sampling system as its predecessor it has several advantages over it, in that it is much more user friendly. The instrument displays zero when the fuel cell is clear and ready to sample, also holds the peak reading, and memorises the last ten peak values which are then easily accessible. The time to peak i.e. the time taken to reach peak reading is usually characteristic of the vapour being analysed. When alcohols such as ethanol, propanol or butanol are present the time to peak is considerably shorter than the time to peak for glutaraldehyde. This enables the instrument to be more selective.




Applications · Specifications

Applications
Glutaraldehyde based solutions are the most commonly used chemical disinfectants in various specialist areas:
  • Medicine
  • Dentistry
  • Veterinary medicine
  • X-ray processing
  • Microbiology
  • Histopathology
  • Paper and leather manufacture
HOSPITAL USE
In a typical hospital application, readings can be taken above and around disinfectant machines, trolley units and in locations such as endoscopy suites, X-ray departments and theatre areas. It can be used to monitor glutaraldehyde when mixing and disposing glutaraldehyde solutions.



Specifications

Sensors

Electrochemical Fuel Cell

Calibration

Glutaraldehyde-Absorbent Standard (Sealed Unit)

Display

3 Digit LCD

Sampling Rate

Dependent on previous reading (0 - 5 minutes)

Sample Volume

10 cm3

Detection Range

0.03 - 4 ppm

Precision

10% at 0.20 ppm

Response Time

Approximately 60 seconds from sampling glutaraldehyde

Battery Life

300 field tests approx. PP3 9v alkaline battery

Weight

Instrument: 270 g
Complete Kit: 750 g

Dimensions

Instrument: 150 x 80 x 34 mm
Carrying Case: 266 x 230 x 50 mm

Kit Includes

Instrument including Battery
1 Calibration Standard
1 Handbook
1 Thermometer

Accessories

Calibration Standard Phenol Filters

Principle of Operation

The glutaraldehyde vapour when drawn across the fuel cell sensor undergoes catalytic oxidation on the platinum surface. This produces an electrical output directly proportional to the glutaraldehyde level in the atmosphere.

HOW TO USE
1. Hold the instrument in the atmosphere to be tested and switch on.
2. Press the sample button
3. The glutaraldehyde level is displayed in approximately 60 seconds.

PHENOL FILTERS
Complete removal of any phenol contaminant is possible with the use of phenol filters.

SELECTIVITY
A high degree of selectivity has been incorporated into the fuel cell sensor. However some substances such as ethanol, propanol and butanol can interfere. By comparing the time to peak with that of the standard it is possible to determine if interfering substances are present and whether more sophisticated analysis is required.

CALIBRATION
The Glutaraldemeter will normally hold calibration for several months. The instrument can however be checked and calibrated in the field, using the PPM Glutaraldehyde standard. The standard consists of glutaraldehyde solution impregnated on a special absorbent material. A headspace sample from this tube can be used to calibrate the instrument.


    Download PDF Download PDF  
    Email Email This Page  
    Save To Favorites Save To Favorites
  

 MORE IMAGES 

· Glutaraldemeter Performance Chart


 DOWNLOADS 

 width= Glutaraldemeter®3 Literature
(PDF, 137KB)


Request Information
Site created by ThomasNet Web Solutions and powered by Navigator Platform
ENMET Corporation | Gas Detection for Health, Safety and Process Monitoring